Voltage and current cannot reach set value

更新时间: Aug 24,2023 Abnormal voltage & current

Solution 1: Current Setting Too Low - Please check if the set current is too low and has gone below the minimum range of the device.

  

Solution 2: Modify the configuration file to suppress prompts for deviations over 0.2% of the range.

Open the BTS 8.0 software, click on the "Settings" in the menu bar, select "System Settings" from the dropdown menu, then choose "Step edit." Make the necessary modifications according to your requirements and save the changes.

 

Solution 3: Check if the fixture is damaged. Verify the proper connections between the fixture, battery, and battery testing equipment.

 

Solution 4: Device calibration (For calibration services, please contact the after-sales department).

 