Discharge curve fluctuations

更新时间: Sep 28,2023 Abnormal curve

Solution 1: Check if it's caused by fixture abnormalities and replace the fixture.


Solution 2: Adjust the sampling frequency of theControl Unit to match that of the Tester.

1. To modify the sampling frequency of the Control Unit, you first need to obtain the Control Unit's IP address.

(1) You can find it on the physical label of the Control Unit.

(2) You can check it on the Control Unit's display screen.

(3) You can find it in the BTS software.

2. Follow these steps:

(1) Open a web browser and enter the Control Unit's IP address: 192.168.10.3.


(2) Access the Neware Equipment Management System and log in with the username "neware" and password "xinwei.

 

(3) Click on the 'Device Configuration' in the left sidebar, where you can modify the Control Unit's sampling frequency.


Note: Ensure that the Control Unit's sampling frequency is consistent with the Tester's sampling frequency.